• Ammonia Analyzer Aztec Aam631
  • Ammonia Analyzer Aztec Aam631
  • Ammonia Analyzer Aztec Aam631
  • Ammonia Analyzer Aztec Aam631
  • Ammonia Analyzer Aztec Aam631
  • Ammonia Analyzer Aztec Aam631

Ammonia Analyzer Aztec Aam631

Display: LED
Appliance: Drinking Water&Wastewater
Portable: Non Portable
Certification: CE
Type: Colorimetric Analyzer
Transport Package: Origial Package
Customization:
Diamond Member Since 2021

Suppliers with verified business licenses

Manufacturer/Factory
  • Overview
  • Description
  • Overview
  • Data
Overview

Basic Info.

Model NO.
AAM631
Trademark
ABB

Product Description

 ABB Ammonia analyzer Aztec AAM631

Description

The Aztec 600 ISE ammonia analyzer provides reliable and accurate measurement of ammonia concentrations for a wide variety of applications including source water monitoring and wastewater nutrient removal.

Ammonia Analyzer Aztec Aam631
Ammonia Analyzer Aztec Aam631
Ammonia Analyzer Aztec Aam631

 

Overview

The Aztec 600 ISE ammonia analyzer uses a robust gas sensing ammonia electrode to provide accurate and reliable continuous measurement of the total ammonia nitrogen (TAN) concentration, that is the sum of the gas NH3 and the cation NH4+.

Featuring ABB's intuitive menu-driven software and the robust ISE sensing system with automatic calibration, users benefit from an analyzer that is both simple-to-operate and maintain.

Users benefits from:
  - Continuous measurement using ISE technology
- Automatic 2-point calibration
- Buffered measurement for maximum accuracy
- Low reagent consumption (up to 3 months continuous measurement before reagent replenishment)
- Temperature-controlled flowcell for measurement stability
- Intuitive software with full colour graphical display
- Data logging and graphical process trending
- Full audit trail capability
- Secure data archiving to SD Card

Data

Measurement range:
0.050 - 1000 ppm NH3 (0.050 - 1000 mg/L)
Measurement Principle:
Gas-sensing ammonia probe
Measurement mode:
Continuous
Accuracy / maximum measured error:
±5 % of reading or ±0.02 ppm for 0.05 to 500 ppm (whichever is the greater)
±7.5 % of reading for 500 to 1000 ppm
Repeatability:
±2 % of reading or ±0.02 ppm (whichever is the greater)
Response time:
Normal operating mode: T90 typically 5 mins
EcoMode: T90 typically <7.5 mins
Resolution:
0.001 ppm for 0 to 10 ppm
0.01 ppm for 10 to 100 ppm
0.1 ppm for 100 to 1000 ppm
Calibration:
2-point, automatic calibration
Measurement units:
mg/l, ppm, ppb, μg/l